Report No 094061R-ITCEP07V06
Page: 117 of 123
Product Notebook
Test Item Voltage dips and interruption
Test Mode Mode 2
Date of Test 2009/04/01 Test Site No.2 Shielded Room
Voltage Dips and
Interruption
Reduction(%)
Angle Test
Duration
(Periods)
Required
Performance
Criteria
Performance
Criteria
Complied To
Test Result
30(161V) 0 25 C A PASS
30(161V) 45 25 C A PASS
30(161V) 90 25 C A PASS
30(161V) 135 25 C A PASS
30(161V) 180 25 C A PASS
30(161V) 225 25 C A PASS
30(161V) 270 25 C A PASS
30(161V) 315 25 C A PASS
>95(0V) 0 0.5 B A PASS
>95(0V) 45 0.5 B A PASS
>95(0V) 90 0.5 B A PASS
>95(0V) 135 0.5 B A PASS
>95(0V) 180 0.5 B A PASS
>95(0V) 225 0.5 B A PASS
>95(0V) 270 0.5 B A PASS
>95(0V) 315 0.5 B A PASS
>95(0V) 0 250 C B PASS
>95(0V) 45 250 C B PASS
>95(0V) 90 250 C B PASS
>95(0V) 135 250 C B PASS
>95(0V) 180 250 C B PASS
>95(0V) 225 250 C B PASS
>95(0V) 270 250 C B PASS
>95(0V) 315 250 C B PASS
Meet criteria A: Operate as intended during and after the test
Meet criteria B: Operate as intended after the test
Meet criteria C: Loss/Error of function
Additional Information
The nominal voltage of EUT is 230V.
EUT stopped operation and could / could not be reset by operator at kV
of Line
.
No false alarms or other malfunctions were observed during or after the test. The
acceptance criteria were met, and the EUT passed the test.