Report No: 094061R-ITCEP07V06
Page: 110 of 123
13. Power Frequency Magnetic Field
13.1. Test Specification
According to Standard : IEC 61000-4-8
13.2. Test Setup
13.3. Limit
Item Environmental
Phenomena
Units Test Specification Performance
Criteria
Enclosure Port
Power-Frequency
Magnetic Field
Hz
A/m (r.m.s.)
50
1
A
13.4. Test Procedure
The EUT and its load are placed on a table which is 0.8 meter above a metal ground
plane measured at least 1m*1m min. The test magnetic field shall be placed at central
of the induction coil.
The test magnetic Field shall be applied 10 minutes by the immersion method to the EUT.
And the induction coil shall be rotated by 90 in order to expose the EUT to the test field
with different orientation (X, Y, Z Orientations).
13.5. Deviation from Test Standard
No deviation.